Calibration & Services

As the manufacturer we maintain laboratory accreditation (NVLAP according to EN17025 and DAkkS according to EN17025) and provide both “mail-in” and “on-site” service options using our modern high precision calibration facilities and high end lab equipment.
We offer complete product support for the life of your Kaye equipment-going beyond typical calibration to provide customers with the highest quality of calibration and repair services. A timely calibration schedule assures accurate and reliable product performance and maximizes equipment uptime and productivity.
Kaye Validator- Factory Calibration
- AS-FOUND, adjustment and AS-LEFT calibration with NIST traceable certification
- SIMs: AS-FOUND, adjustment and ASLEFT calibration of the internal cold junction references (V2K calibration includes the calibration of up to 3 Sims)
- Very low measurement uncertainties for calibration to guarantee high performance
- Includes 8 hour elevated temperature soak to identify potential electronic latent defects
- Free clock battery replacement included
- Power loss test to check the back-up battery. If required a free exchange of backup battery
|
 |
Kaye IRTD-400 - Intelligent Temperature Standard - Factory Calibration
- NVLAP AS-LEFT certificate includes -196°C, 0°C, 100°C and 400°C points
- NVLAP AS-FOUND certificate includes -196°C, 0°C, 100°C, 165°C and 400°C points
- Annealing Process at 400°C for a minimum of 15 hours to release sensor element stress and to offer more stability and improved performance between calibration cycles
- We always adjust and reprogram the IRTD internal calibration coefficients over the full range
|
 |
Kaye Valprobe™ Wireless Process Validation and Monitoring - Factory Calibration
- NVLAP AS-FOUND calibration at minimum 4 points for Temperature, Humidity & Pressure
- High accuracy NVLAP AS-LEFT calibration at minimum 4 points
- Adjustment of internal coefficients to achieve optimal reading
- Includes free replacement of Lithium batteries and o-ring gaskets
|
 |
Kaye RF ValProbe™ Wireless Thermal Validation System - Factory Calibration
- High accuracy NVLAP AS-LEFT calibration at minimum 4 points for Temperature and Humidity
- NVLAP AS-FOUND calibration at minimum 4 points
- Adjustment of internal coefficients to achieve optimal reading
- Includes free replacement of Lithium batteries
|
 |
Kaye LTR/HTR Thermal References - Factory Calibration
- AS-FOUND calibration at 50, 100, 150, 200, 250, 300, 350 and 400°C with HTR
- Adjustment of internal calibration coefficients
- AS-LEFT calibration at 50, 100, 150, 200, 250, 300, 350 and 400°C with HTR
- Cycle and stability test at highest temperature, Cut-Out Safety Test
- Free firmware upgrade upon request
|
 |
Kaye Validator - Factory Calibration
- AS-FOUND, adjustment and AS-LEFT calibration with DKD traceable certification
- SIMs: AS-FOUND, adjustment and ASLEFT calibration of the internal cold junction references (V2K calibration includes the calibration of up to 3 Sims)
- Very low measurement uncertainties for calibration to guarantee high performance
- Free clock battery replacement included
- Verification of all 36 channels and communication ports
- Internal clock verification
- Power loss test to check the back-up battery. If required a free exchange of backup battery
|
 |
Kaye IRTD-400 - Intelligent Temperature Standard - Factory Calibration
- DKD AS-LEFT certificate includes -196°C, -60°C, 0°C, 100°C, 260°C and 420°C points
- DKD AS-FOUND certificate includes -196°C, -60°C, 0°C, 100°C, 260°C and 420°C points
- Annealing Process at 420°C for a minimum of 70 hours to release sensor element stress and to offer more stability and improved performance between calibration cycles
- We always adjust and reprogram the IRTD internal calibration coefficients over the full range to achieve optimal reading
|
 |
Kaye Valprobe™ Wireless Process Validation and Monitoring - Factory Calibration
- DKD AS-FOUND calibration at minimum 4 points for Temperature, Humidity & Pressure
- High accuracy DKD AS-LEFT calibration at minimum 4 points
- Adjustment of internal coefficients to achieve optimal reading
- Includes free replacement of Lithium batteries and o-ring gaskets
- Logger maintenance, electrical tests
|
 |
Kaye RF ValProbe™ Wireless Thermal Validation System - Factory Calibration
- High accuracy DKD AS-LEFT calibration at minimum 4 points for Temperature and Humidity
- DKD AS-FOUND calibration at minimum 4 points
- Adjustment of internal coefficients to achieve optimal reading
- Includes free replacement of Lithium batteries
|
 |
Kaye LTR/HTR Thermal References - Factory Calibration
- AS-FOUND calibration at 50, 100, 150, 200, 250, 300, 350 and 400°C with HTR
- Adjustment of internal calibration coefficients
- AS-LEFT calibration at 50, 100, 150, 200, 250, 300, 350 and 400°C with HTR
- Cycle and stability test at highest temperature, Cut-Out Safety Test
- Well to Well uniformity test at 121°C
|
 |