Kaye HTR & LTR Series Uniform Dry Wells

NEW ! LTR-90

 

The HTR and LTR Series temperature references are dry wells specifically designed to fully automate sensor calibration when used with GE Kaye validation software and traceable temperature standards.

 

Kaye dry wells employ unique inserts which minimize cooling of the thermocouple tips due to heat conduction providing an uncertainty of merely 0.1°C. Without proper inserts, for example, transfer uncertainty in excess of 0.5°C can occur with 22 gage, type T thermocouples.

 

Both the HTR and LTR series models feature fast heat up and cool down, a capacity to hold from 18 to 24 thermocouples and require no messy oils.

 

Calibrating sensors in the environment where they will be used provides the highest degree of accuracy as system temperature coefficient errors are eliminated. The references are lightweight and rugged so you can calibrate sensors almost anywhere.

 

The HTR-400 with an operating range from 25°C above ambient to 400°C is ideal for calibrating sensors for autoclaves, dry heat ovens and sterilization tunnels. For freezers, cold rooms, incubators, and autoclaves, select from the LTR series of dry wells with operating ranges of -95 to 140°C or -25 to 140°C.

Both the HTR and LTR series models feature fast heat up and cool down, a capacity to hold from 18 to 24 thermocouples and require no messy oils.

Calibrating sensors in the environment where they will be used provides the highest degree of accuracy as system temperature coefficient errors are eliminated. The references are lightweight and rugged so you can calibrate sensors almost anywhere.

The HTR-400 with an operating range from 25°C above ambient to 400°C is ideal for calibrating sensors for autoclaves, dry heat ovens and sterilization tunnels.

For freezers, cold rooms, incubators, and autoclaves, select from the LTR series of dry wells with operating ranges of -95 to 140°C or -25 to 140°C.

Features and Benefits for LTR90

  • –95 ºC to 140 ºC Ultra-Cool Dry Block Calibrator
  • Stirling cooler technology: reaches –80 ºC in 60 min; –95 ºC in 80 minutes
  • Temperature stability of ±0.015 ºC
  • Actual Uniformity 0.05°C Full Range
  • Software interface with Kaye Validator for automatic sensor calibration utility
  • Custom block design to accommodate 2 reference probes (IRTD) and four inserts with three thermocouples per insert
  • Conformance to Euramet/cg-13
  • Ergonomics/industrial design with two carrying handles for portability
  • Rubber insulator cap to prevent frost buildup
  • Weight: 16 kg (35 lbs.)

 

The HTR 400 is ideal for high-temperature applications such as autoclaves, dry heat ovens and sterilizer tunnels.

The LTR models offer low-temperature performance for applications including freezers, cold rooms, incubators and autoclaves. The LTR model selection should be based on the application’s low-temperature point.

 

Temperature Range at 23°C

-95° to 140°C (-139°F to 284°F)

 

Display Accuracy

±0.2°C Full Range

 

Resolution

0.01°C

 

Accuracy with External Reference

± 0.05 °C Full Range

 

Stability

±0.015°C Full Range

 

Stabilization Time

15 min.

 

Axial Uniformity at 40 mm (1.6 in.)

±0.05°C Full Range

 

Radial Gradient

±0.01°C Full Range

 

Loading Effect (with a 6.35 mm Reference Probe and Three 6.35 mm Probes)

±0.006C Full Range

 

Loading Effect (versus Display with 6.35 mm Probes)

±0.25°C at -95°C
±0.10°C at 140°C

 

Operating Conditions

0°C to 35°C
0% to 90% RH (non-condensing)

 

Environmental Conditions for All Specifications(except Temperature Range)

13°C to 33°C

 

Immersion (Well) Depth

160 mm (6.3 in.)

 

Insertion Holes

2 reference holes (IRTD)
and 4 T/C holes (3 T/C/hole).

 

Heating Time

-95°C to 140°C: 40 min.

 

Cooling Time

23°C to -90°C: 80 min.
23°C to -95°C: 90 min.
140°C to 23°C: 60 min.