UPS II Smart Loop Calibrator for Milliamps

UPS II Loop Calibrator

 

The UPS-II loop calibrator is an advanced handheld process calibrator tool. It can power (24 VDC) and read 2-wire transmitters to perform field calibrations. Output currents are adjustable to a resolution of 10 microamps in both source and transmitter simulation.

 

 

 

 

 

 

 

 

 

 

 

 


UPS II Specifications

  • Sources and reads milliamps
  • Powers and reads 2-wire transmitters
  • Reads and simulates DP transmitters in % of flow
  • Auto-ramp cycling for endurance tests
  • Auto-stepping for hands-free calibration
  • Fixed calibration currents
  • Fixed currents for valve stroking
Output Range 0 to 21.00 mA
Input Range 0 to 21.00 mA
Accuracy 0.05% of range
Resolution 10 microamps
Temperature Effect 0.0015% /°F (0.003% /°C)
Internal Loop-Power 24 VDC stabilized
External Loop-Power 56 VDC maximum
Measurement Input 12 O including fuse
Drive Capability 900 O
Operating Temperature 15°F to 120°F (-10°C to 50°C)
Battery Life Sources 12 mA: 13 hours
Measures: 64 hours
Batteries 4x 1.5V AA Batteries
Loop Mismatch Flashes "loop"
Low Battery Flashes "lobat"
Over range Flashes "EEEE"
Reversed Polarity Flashes "pol"
Housing High impact ABS
Size 3 in x 5 in x 0.83 in (77 mm x 124 mm x 21 mm)
Weight 9 oz (285 g) including batteries
Fixed Steps The calibrator can source or simulate fixed calibration currents to test or calibrate receivers and to stroke valves. Different current levels are available from the menu.

4 to 20 mA linear 4, 8, 12, 16, 20 mA
0 to 20 mA linear 0, 5, 10, 15, 20 mA
4 to 20 mA sq.rt (flow) 4, 5, 8, 13, 20 mA
0 to 20 mA sq.rt (flow) 0, 1.25, 5, 11.25, 20 mA
4 to 20 mA 3.8, 4, 4.2, 12, 19, 20, 21 mA

Output currents can be changed manually or automatically. The time between steps is 10 seconds in the auto-stepping mode.
In the valve mode, auto-stepping will be: 3.8, 4, 4.2 or 19, 20, 21 mA
Ramping The calibrator can source or simulate with high resolution a continuous up/down cycle between 4 and 20 mA or 0 and 20 mA. The travel time between limits is 60 seconds.
Reading in Percentage The calibrator can display readings in percentage of a 4 to 20 mA or 0 to 20 mA span and can also read or simulate dP-transmitters directly in percentage of flow.

Descriptions English
UPS II Brochure UPS II Brochure