| Fluid Types |
Acoustically conductive fluids, including most clean liquids, and many liquids with entrained solids or gas bubbles. Maximum void fraction depends on transducer, interrogation carrier frequency, path length and pipe configuration. |
| Pipe Sizes |
- Clamp-on transducers: 0.5 to 300 in. (12.7 mm to 7.6m) and larger
- Wetted transducers: 1 in to 200 in (25.4 mm to 5 m) and larger
|
| Pipe-Wall Thickness |
Up to 3 in (76.2 mm) |
| Pipe Materials |
All metals and most plastics. Consult GE for concrete, composite materials, and highly corroded or lined pipes. |
| Flow Accuracy (Velocity) |
0.5% of reading (achievable with process calibration) |
| Typical Clamp-On Flow Accuracy (Velocity) |
- Pipe ID>6 in (150 mm): ±1% to 2% of reading
- Pipe ID<6 in (150 mm): ±2% to 5% of reading
|
| Typical Wetted Flow Accuracy (Velocity) |
±1% of reading accuracy depends on pipe size and installation and whether measurement is one-path or two path. |
| Repeatability |
±0.1% to 0.3% of reading |
| Range (Bidirectional) |
–40 to 40 ft/s (–12.2 to 12.2 m/s) |
| Rangeability (Overall) |
400:1 |
| Specifications assume a fully developed flow profile (typically 10 diameters upstream and 5 diameters downstream of straight pipe run) and flow velocity greater than 1 ft/s (0.3 m/s). |
| Measurement Parameters |
Volumetric flow, totalized flow and flow velocity |
Electronics
|
| Flow Measurement |
Patented Correlation Transit-Time mode |
| Enclosure |
Epoxy-coated aluminum weatherproof Type 4X/IP66 |
| Dimensions |
Standard: Weight 2 lb (0.9 kg), |
| size (h x w x d) |
7.25 in x 5.9 in x 3.5 in (184 mm x 150 mm x 89 mm) |
| Channels |
- Standard: One channel
- Optional: Two channels (for two pipes or two-path averaging)
|
| Display |
2-line x 16 character backlit LCD display, configurable to display up to four measurement parameters in sequence |
| Keypad |
Six-button internal keypad |
| Power Supplies |
- Standard: 85 to 265 VAC, 50/60 Hz
- Optional: 12 to 28 VDC, ± 5%
|
| Power Consumption |
20W maximum |
| Operating Temperature |
14°F to 131°F (–10°C to 55°C) |
| Storage Temperature |
–40°F to 158°F (–40°C to 70°C) |
| Standard Inputs/Outputs |
- One 0/4 to 20 mA isolated output per channel, 600 Ω maximum load
- One frequency/pulse rate/totalizer output per channel, optically isolated, 3A maximum, 100 VDC maximum, 1W maximum, from 0.1 to 10 kHz
|
| Digital Interfaces |
- Standard: RS232
- Optional: RS485 (multiuser)
|
| European Compliance |
System complies with EMC Directive 89/336/EEC, 73/23/EEC LVD (Installation Category II, Pollution Degree 2) and transducers comply with PED 97/23/EC for DN<25 |
Clamp-On Ultrasonic Flow Transducers
|
| Temperature Ranges |
- Standard: -40° to 150°C (-40° to 302°F)
- Optional: -200° to 400°C (-328C to 752°F)
See specific transducer for exact temperature range.
|
| Mountings |
Stainless steel chain or strap, welded or magnetic clamping fixtures |
| Area Classifications |
- Standard: General purpose
- Optional: Weatherproof Type 4/IP65
- Optional: CSA certified (Class I, Division 1, Groups B, C and D)
- Optional: ATEX certified (ll 2 GD Eex md llC T6 T=80°C)
- Optional: Submersible IP67/68
|
Wetted Ultrasonic Flow Transducers
|
| Temperature Range |
–40°F to 212°F (–40°C to 100°C) |
| Pressure Range |
0 to 3000 psig (1 to 207 bar) |
| Materials |
- Standard: Stainless steel
- Optional (for Pan-Adapta® Plugs): Titanium, Hastelloy® alloy, Monel® alloy, duplex, CPVC, PVDF and others Pan-Adapta plugs allow installation and removal of wetted transducers without interrupting the process or emptying the pipe.
|
| Process Connections |
- Standard: 1 in or 3/8 in NPTM
- Optional: RF flanged, socket weld, fuse bond and others
|
| Mountings |
Flanged flowcell, hot tap or cold tap |
| Area Classifications |
- Standard: General purpose
- Optional: Weatherproof Type 4/IP65 submersible
- Optional: Submersible IP67/68
|
| Transducer Cables |
• Standard: One pair of coaxial cables, type RG62 AU, or as specified for transducer type • Optional: Lengths up to 1000 ft (330 m) maximum |
| Additional Options |
PanaView™ PC-Interface Software The AquaTrans AT868 communicates with a PC through a serial interface and Windows® operating systems. Consult the manual for details on sites, logs, and other operations with a PC. |