The Moisture Image Series 1 is the multichannel, multifunction top-of-the-line model in the GE moisture series analyzer family. The Moisture Image Series 1 Analyzer measures moisture, pressure and temperature in non-aqueous liquids and gases. There is also an option for measuring oxygen concentration in gases. In addition, the auxiliary inputs can accept readings from any sensor with 0/4 to 20 mA or 0 to 4 V output, including a variety of GE process control instruments. This feature makes the Moisture Image Series 1 a true multifunction analyzer, providing cost savings through system integration.

 

For additional cost savings, the Moisture Image Series 1 Analyzer is available with up to six channels to significantly reduce the cost-per-measurement point while providing a variety of configurations to meet the needs of specific applications.

 

This state-of-the-art instrument features simultaneous real-time display of six points of moisture, oxygen, temperature, pressure and other data from up to six channels on a large, backlit liquid crystal display (LCD). Menu-driven function keys provide easy access to data displayed in matrix and line-graph formats.

 

The Moisture Image Series 1 Analyzer is complemented by the Moisture Image Series probe. It measures moisture, temperature and pressure, eliminating the need for additional external pressure transmitters and related power supplies. It is suitable for laboratory and industrial moisture measurement applications in gases and non-aqueous liquids over a wide range of process conditions. The Moisture Image Series 1 can also be used with any Panametrics moisture probe configuration. All moisture probe calibrations are NIST traceable.

Moisture Image Series 1 Specifications

  • Optional pressure and temperature inputs used to calculate various parameters
  • Calibrations traceable to National Institute of Standards and Technology (NIST)
  • Rack, bench, panel, weatherproof (Type 4X) and explosion-proof (Type 7) configurations available
  • Computer-enhanced response software for abrupt dry-down applications
  • Real time matrix display of six parameters simultaneously
  • Measures other parameters via auxiliary inputs
  • Displays trace measurements in ppbv
  • Built-in data logging
  • PCMCIA card slot for additional memory and instrument program updates

This multichannel analyzer measures moisture in gases and non-aqueous liquids and oxygen (optional) in gases. Designed for permanent installations, it is used in conjunction with Moisture Image Series, TF and M Series moisture probes for industries including:

  • Petrochemical
  • Natural gas
  • Industrial gas
  • Semiconductor
  • Furnace gas/heat treating
  • Power generation
  • Air dryer
  • Pharmaceutical
  • Aerospace

Electronics

Intrinsic Safety Galvanic isolation and energy-limiting circuitry for all inputs (except auxiliary):
BAS01ATEX7097 II (1) G [EEx ia] IIC –20°C to +50°C;
CSA C US Class I, Division 2, Groups B,C&D T4A–Assoc.
Elec. Apparatus [Exia] and Class I, Division 1,Groups A,B,C&D T4
European Compliance Complies with EMC Directive 89/336/EEC and 73/23/EEC LVD (Installation Category II, Pollution Degree 2)
Inputs
  • Moisture: One to six channels
  • Temperature: One to six channels
  • Pressure: One to six channels
  • Oxygen: One to six channels • Auxiliary (optional): Two per installed channel; may be used for 0/4 to 20 mA and 0 to 4 V devices
Analog Outputs Two per channel; internally optically isolated; 10-bit (0.1%) resolution
Programmable Switch-Selectable Outputs
  • 0 to 2 V, 10 S minimum load resistance
  • 0/4 to 20 mA, 400 S maximum series resistance
Digital Outputs RS232 serial communications port; information is transmitted as ASCII
A/D Resolution 12-bit or 16-bit with MIS probe
Computer-Enhanced Response Software Standard: Provides fast measurements in abrupt dry-down applications (three- to five-minute speed of response)
Data Logging Up to 12 parameters simultaneously; stored in RAM; log length depends on number of items and time interval
Memory
  • Standard: Battery-backed 64 kB RAM 
  • Optional: Additional 1 MB (or more) with PCMCIA card
PC-Interface Software Optional: PanaView™ instrument interface software
Alarm Relays
  • Two optional Form C relays per channel SPDT; rated for 2 A at 28 VAC/28 VDC
  • Standard and hermetically sealed (for Division 2 hazardous areas per CSA) available for high and low limits on each channel; set to any level within the range of the instrument
Alarm Set Point Accuracy ±0.1°C dew point
Display Graphic/text backlit LCD
Display Functions Displays up to six channel/parameter combinations simultaneously using text and/or graphics
Power Requirements Universal power supply adjusts automatically for 90 to 260 VAC, 50/60 Hz
Temperature
  • Operating: 32°F to 140°F (0°C to 60°C)
  • Storage: –22°F to 158°F (–30°C to 70°C)
Data Retention Calibration data stored in battery-backed RAM
Autocalibration Occurs on power-up and at user-selected time intervals
Configuration Dimensions (h x w x d)
  • Rack mount: 5.22 x 19.00 x 17.03 in (13.26 cm x 48.26 cm x 43.26 cm)
  • Bench mount: 5.85 in x 13.94 in x 17.03 in (14.86 cm x 35.41 cm x 43.26 cm)
  • Panel mount: 8.25 in x 16.50 in x 17.03 in (20.96 cm x 41.91 cm x 43.26 cm)
  • Weatherproof: 22.23 in x 16.23 in x 9.48 in (56.46 cm x 41.22 cm x 24.08 cm)
  • Explosion-proof: Consult factory

Moisture Measurement

Type GE Moisture Image Series, TF, and M Series thin-film aluminum oxide probes
Calibration Ranges (Dew/Frost Point)
  • Standard: 68°F to -112°F (20°C to –80°C) with data to -166°F (–110°C)
  • Ultralow: -58°F to -148°F (–50° to –100°C) with data to -166°F (–110°C)
  • Extended high: 140°F to -112°F (60°C to –80°C) with data to -166°F (–110°C)
Accuracy (Dew/Frost Point)
  • ±1.1°F (±2°C) from 140°F to -85°F (60°C to –65°C)
  • ±1.6°F (±3°C) from -87°F to -166°F (–66°C to –110°C)
Repeatability (Dew/Frost Point)
  • ±0.3°F (±0.5°C) from 140°F to -85°F (60°C to –65°C)
  • ±0.5°F (±1.0°C) from -87°F to -166°F (–66°C to –110°C)
Operating Pressure 5 of Hg to 5000 psig (345 bar) limited by optional pressure sensor—see pressure sensor ranges
Other Moisture Parameters Relative humidity (RH), parts per million by volume (ppmv), parts per million by weight (ppmw), parts per billion by volume (ppbv), pounds per million standard cubic feet (lb/MMSCF)

Temperature Measurement

Type Optional thermistor built into moisture probe
Range –22°F to 158°F (–30°C to 70°C)
Accuracy ±0.3°F (±0.5°C) at –22°F (–30°C)

Pressure Measurement

Type Optional transducer built into TF and Moisture Image Series moisture probes or standard external pressure transmitter, 24 V, 4 to 20 mA
Range 0 to 5000 psig (1 to 345 bar); specify type and pressure range
Accuracy ±1% of full scale
Pressure Rating Three times the span of the available range to a maximum of 7500 psig (518 bar)

 

Oxygen Measurement


Optional nondepleting electrolytic oxygen sensor.
Additional details and specifications are available on request.
Descriptions Chinese Dutch English Français Deutsch Italiano Japanese Polish Português Russian Español
MIS 1 Bench-Mount Package Drawing 
MIS 1 Bench-Mount Package Drawing
MIS 1 Panel-Mount Package Drawing 
MIS 1 Panel-Mount Package Drawing
MMS 3 Weatherproof Package Drawing 
MMS 3 Weatherproof Package Drawing
BASEEFA System Drawing for MIS 1 IS System BASEEFA System Drawing for MIS 1 IS System BASEEFA System Drawing for MIS 1 IS System BASEEFA System Drawing for MIS 1 IS System
BASEEFA System Drawing for MMS 3 IS System BASEEFA System Drawing for MMS 3 IS System
Basic Hygrometry Principles Basic Hygrometry Principles
CSA-NRTL System Drawing for MIS 1 IS System CSA-NRTL System Drawing for MIS 1 IS System
M Series Moisture Probe M Series Moisture Probe M Series Moisture Probe M Series Moisture Probe M Series Moisture Probe M Series Moisture Probe
MIS 1 Bench-Mount Package Drawing MIS 1 Bench-Mount Package Drawing MIS 1 Bench-Mount Package Drawing
MIS 1 Interconnection Drawing for Rack, Bench & Panel MIS 1 Interconnection Drawing for Rack, Bench & Panel
MIS 1 Panel-Mount Package Drawing MIS 1 Panel-Mount Package Drawing MIS 1 Panel-Mount Package Drawing
MIS 1 Rack-Mount Package Drawing MIS 1 Rack-Mount Package Drawing MIS 1 Rack-Mount Package Drawing
MIS 1 Service Manual MIS 1 Service Manual
MIS 1 Startup Guide MIS 1 Startup Guide
MIS 1 Weatherproof Package Drawing MIS 1 Weatherproof Package Drawing
MIS Probe 2 Aluminum Oxide Moisture Probe Brochure MIS Probe 2 Aluminum Oxide Moisture Probe Brochure
MIS Probe Datasheet MIS Probe Datasheet MIS Probe Datasheet MIS Probe Datasheet MIS Probe Datasheet
MIS1 (w/703-1250 Controller Board) MIS1 (w/703-1250 Controller Board)
MIS1 (w/703-1250 Controller Board) Service MIS1 (w/703-1250 Controller Board) Service
MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual MIS1 Hygrometer Operating Manual
MIS1 Quick Start MIS1 Quick Start
MMS 3 Bench-Mount Package Drawing MMS 3 Bench-Mount Package Drawing
MMS 3 Interconnection Drawing for Rack, Bench & Panel MMS 3 Interconnection Drawing for Rack, Bench & Panel
MMS 3 Interconnection Drawing for Weatherproof Package MMS 3 Interconnection Drawing for Weatherproof Package
MMS 3 Panel-Mount Package Drawing MMS 3 Panel-Mount Package Drawing
MMS 3 Rack-Mount Package Drawing MMS 3 Rack-Mount Package Drawing
Moisture Image Series 1 Analisador de Umidade da Panametrics Moisture Image Series 1 Analisador de Umidade da Panametrics
Moisture Image Series 1 Analizzatore di umidita Panametrics Moisture Image Series 1 Analizzatore di umidita Panametrics
Moisture Image Series 1 Brochure Moisture Image Series 1 Brochure Moisture Image Series 1 Brochure Moisture Image Series 1 Brochure
Moisture Probe Calibration Moisture Probe Calibration Moisture Probe Calibration Moisture Probe Calibration
Moisture Series Moisture Series
Nondepleting Oxygen Cell Nondepleting Oxygen Cell Nondepleting Oxygen Cell Nondepleting Oxygen Cell Nondepleting Oxygen Cell
Three-Function Moisture Probe Three-Function Moisture Probe Three-Function Moisture Probe Three-Function Moisture Probe Three-Function Moisture Probe