DPI 620 Advanced Modular Calibration and HART® Communication System

Now ATEX Certified!
The DPI 620 Series an ultra-compact electrical, temperature and frequency calibrator and HART communicator, which provides simultaneous measurement and source capabilities with a fully featured HART digital interface. It integrates with pressure modules and pressure generation stations to form uniquely capable pressure test and calibration systems.
The Intrinsically Safe version of the DPI 620 Series is now available allowing configuration and multi-function calibration in hazardous areas. The ATEX and IECEx certification allows the instrument to be employed globally in Zone 0 areas. The certification also applies to the instrument’s battery pack, permitting hot swapping within hazardous locations. The approval now further extends the range of potential applications of the powerful communicator/calibrator, especially in the oil and gas and process sectors. As with all instruments in the DPI 620 family, the IS version is rugged and compact and weather-proofed to IP65.
Features
Flexible and Modular SystemIts basic component is an ultra-compact, electrical, frequency and temperature calibrator and HART communicator, which provides simultaneous measurement and source capabilities with a fully featured HART digital interface. Pressure measurement is provided by interchangeable pressure modules that can be attached to the calibrator by a pressure module carrier. If pressure generation is required the calibrator and pressure module can be attached to one of three pressure generation stations, to form a fully-integrated pressure calibrator, of unparalleled performance. The simple yet sophisticated design and concept combines, for the first time, an advanced electrical calibrator with state-of-the-art pressure measurement and generation, so that there is no longer any need to compromise one measurand capability in favour of another.
Compact and PowerfulThe DPI 620 electrical calibrator can measure and source mA, mV, V, Ohms, frequency and a variety of RTDs and T/Cs. It provides an isolated 24 V loop power supply to energise devices and control loops and a stabilized dc voltage supply for ratio metric transducers. The high resolution, color touch screen of the calibrator shows only those buttons necessary for the immediate, selected tasks. The buttons have been sized so that they can be operated by a gloved hand, with no need for a stylus. The display can be configured to show up to six reading windows and can show electrical and temperature inputs and outputs, two pressure module readings, an IDOS pressure measurement, and a USB-connected device, as well as a HART summary window or full HART communicator with active mA measurement and actual primary variable readings.
The reading windows also act as active keys and, when pressed, they expand to show more detailed information or to provide advance functions such as step and ramp outputs. color is used to emphasize important information, for example, red for critical or failed values and green for pass or within tolerance.
The DPI 620 includes comprehensive documenting features for automating instrument calibration when used in conjunction with Intecal Calibration Management software. The compact and lightweight calibrator is designed for rugged field application, with a robust casing and weather-proofing to IP 65. It features a high-speed digital interface for connecting to pressure modules and pressure generating stations. This interface will also allow expansion to meet future requirements when new modules and stations become available. It also features USB connectivity and a microSD card slot with 2GB of memory installed as standard. (The IS version has 1GB of fixed internal memory.)
|
High-performance HART CommunicatorThe DPI 620 multifunction calibrator now incorporates a fully featured HART® communicator which surpasses market leading communicators in functionality, ease of use and upgrade support such that a single hand held device can now configure the full range of HART devices and perform mA trims and sensor calibrations without the need for secondary equipment such as ammeters, calibrators, power supplies and loop resistors. This can significantly reduce equipment inventory, cost of ownership and greatly simplify maintenance activities The DPI 620 contains the latest application software and a complete library of registered HART device descriptions. Upgrades can be simply downloaded free of charge from our website.
MultifunctionThe Windows CE-driven version of the electrical calibrator provides all the computing power of a conventional hand-held PC or PDA. It provides standard Windows file management and offers popular Windows features to allow users to create text documents and view common file types including Excel, Power Point, Word, pdf and images. This means that operators can consult user manuals, training presentations, data sheets, installation drawings etc while in the field or on the plant.
ATEX and IEC Approved
The four basic system components are available with ATEX and IEC approval for use in hazardous areas. This provides a flexible and expandable system that can be configured and operated safely in areas with potentially explosive atmospheres.
DPI 620 CE with WiFi*The DPI 620 CE version can be further enhance with wireless IEEE 802.11g communications. For the first time, in a calibrator of this type, it is possible to link to the Internet and remote networks in order to access information and to transfer data. This powerful feature will benefit service technicians who spend extended periods away from head office and for those who need instant access to data, safety information, system drawings, product datasheets etc., while on the move. It will also provide an interface to future system modules when a physical connection is a hindrance. *Not available with IS version. |
Applications
Wide application coverage
- Instrumentation installation, commissioning, maintenance and calibration
- HART device configuration, trimming and calibration
- Integrates with leading calibration software solutions
- PDA for documenting routine maintenance and viewing Word, Excel, Power Point, PDF and images
- System measurement and monitoring
- Indicator, recorder and controller testing
- Process loop set up and diagnostics
- Switch, trip and safety system testing
- Multi-channel data logging for system testing and diagnostics
Multiple measure and source parameters for wide application coverage

Multiple reading capabilities essential for instrument calibration and loop set-up

Specifications
DPI 620/CE
| Measure | mA, V DC, V AC, RTDs, TCs, Ohms, Freq., Pulses |
|---|---|
| Source/simulate | mA, V DC, RTDs, TCs, Ohms, Freq., Pulses |
| Accuracy | From 0.0025%rdg + 0.002%FS |
| Loop power | 24V |
| Display | Size: 110 mm (4.3 in) diagonal; 480 x 272 pixels LCD: Colour display with touch-screen |
| Languages | English {Default}, Chinese, French, German, Italian, Portuguese, Russian, Spanish |
| Operating temperature | -10° to 50°C (14° to 122°F) |
| Storage temperature | -20° to 70°C (-4° to 158°F) |
| Ingress Protection | IP65 (Dust-tight, jets of water) |
| Humidity | 0 to 90% RH Non condensing |
| Shock / Vibration | BS EN 61010:2001; Def Stan 66-31, 8.4 cat III, 1 m Drop Tested |
| EMC | Electromagnetic compatibility: BS EN 61326-1:2006 |
| Electrical safety | Electrical – BS EN 61010 : 2001 |
| Pressure safety | Pressure Equipment Directive - Class: Sound Engineering Practice (SEP) |
| Certification | CE |
| Size (L: W: H) | DPI 620 only: 183 x114 x 42 mm (7.2 x 4.5 x 1.7 in) + MC 620: ≈ 265 x 114 x 64 mm ( 10.4 x 4.5 x 2.5 in) + PM 620: ≈ 265 x 114 x 93 mm (10.4 x 4.5 x 3.7 in) |
| Weight | DPI 620 only: ≈ 575 g (1.3 lb) – battery included. MC 620 only: ≈ 640 g (1.4 lb). PM 620 only: ≈ 100 g (0.2 lb). |
| Power supply | Lithium-Polymer battery (GE Part number : IO620-Battery); Capacity: 5040 mAh (minimum), 5280 mAh (typical); Nominal voltage: 3.7 V. Charge temperature: 0° to 40°C (32° to 104°F) Discharge temperature: -20° to 60°C (-4° to 140°F). Note: For best battery performance, keep the temperature less than 60°C (140°F). Charge/discharge cycles: > 500 > 70% capacity. |
| Duration | Measure functions (CH1): ≈ 12 hours continuous. Dual Function, mA measure (CH2): ≈ 7 hours (24 V Source at 12 mA) |
DPI 620 IS and DPI 620 IS CE (INTRINSICALLY SAFE VERSION)
| Measure | mA, V DC, RTDs, TCs, Ohms, Freq., Pulses |
|---|---|
| Source/simulate | mA, V DC, RTDs, TCs, Ohms, Freq., Pulses |
| Accuracy | From 0.0025%rdg + 0.002%FS |
| Loop power | 24V |
| Display | Size: 110 mm (4.3 in) diagonal; 480 x 272 pixels OLED Colour display with touch-screen |
| Languages | English {Default}, Chinese, French, German, Italian, Portuguese, Russian, Spanish |
| Operating temperature | -10° to 40°C (14° to 104°F) |
| Storage temperature | -20° to 70°C (-4° to 158°F) |
| Ingress Protection | IP65 (Dust-tight, jets of water) |
| Humidity | 0 to 90% RH Non condensing |
| Shock / Vibration | BS EN 61010:2001; Def Stan 66-31, 8.4 cat III, 1 m Drop Tested |
| EMC | Electromagnetic compatibility: BS EN 61326-1:2006 |
| Electrical safety | Electrical – BS EN 61010 : 2001 |
| Pressure safety | Pressure Equipment Directive - Class: Sound Engineering Practice (SEP) |
| Certification | CE |
| Size (L: W: H) | DPI 620 only: 183 x114 x 55 mm (7.2 x 4.5 x 2.2 in) + MC 620: ≈ 265 x 114 x 77 mm ( 10.4 x 4.5 x 3 in) + PM 620: ≈ 265 x 114 x 93 mm (10.4 x 4.5 x 3.7 in) |
| Weight | DPI 620 only: ≈ 1.1 kg (2.4 lb) – battery included. MC 620 only: ≈ 800 g (1.8 lb). PM 620 only: ≈ 100 g (0.2 lb). |
| Power supply | NiMH battery (GE Part number: IO620IS-Battery); Capacity: 4000 mAh (typical); Nominal voltage: 3.6 V. Charge temperature: 0° to 40°C (32° to 104°F) Discharge temperature: -10° to 40°C (-14° to 104°F). Note: For best battery performance, keep the temperature less than 60°C (140°F). Charge/discharge cycles: > 500 > 70% capacity. |
| Duration | Measure functions (CH1): ≈ 8 hours continuous. Dual Function, mA measure (CH2): ≈ 7 hours (24 V Source at 12 mA) |
| Approval | Baseefa10ATEX0010X IECEx BAS 10.0002X Ex II 1 G Ex ia IIC T4 Ga (-10°C ≤ Ta ≤ +40°C) EN60079-0:2009 Electrical apparatus for Potentially |
| EN60079-0:2009 | Electrical apparatus for Potentially Explosive Atmospheres – General Requirements. (Harmonized) (IEC 60079-0:2007 Edition 5) |
| EN60079-11:2007 | Electrical apparatus for Potentially Explosive Atmospheres – Intrinsic Safety ‘i’. (Harmonized) (IEC 60079-11:2006 Ed 5) |
Pressure measurement
Pressure module PM 620
| Ranges | from 25 mbar / 10 inH2O to 1,000 bar / 15,000 psi |
|---|---|
| Accuracy | from 0.005%FS |
| Maximum intermittent pressure | 2 x FS |
| Maximum working pressure | 110% FS |
| Sealing | IP 65 (protected against dust and jets of water) |
| Operating temperature | -10 to 50°C (14 to 122 °F) |
| Storage temperature | -20 to 70 °C (-4 to 158 °F) |
| Humidity | 0 to 90% RH non condensing |
| Shock and vibration | BS En 61010:2001;Def stan 66-31, 8.4 cat III), 1m Drop Tested |
| EMC | BS EN 61326-1:2006 |
| Electrical safety | BS EN 61010:2001 |
| Pressure safety | Pressure equipment directive class SEP |
| Approval | CE marked |
| Size and weight | L.56mm, Dia.44mm, 106g maximum |
Pressure module PM 620 IS (intrinsically safe version)
| Ranges | from 25 mbar / 10 inH2O to 1,000 bar / 15,000 psi |
|---|---|
| Accuracy | from 0.005%FS |
| Operating temperature | -10 to 40°C(14 to 104 °F) |
| Approval | Baseefa10ATEX0012X IECEx Bas 10.0004X Ex II 1G Ex ia IIC T4 Ga(-10°C ≤ Ta ≤ +50°C) |
| EN60079-0:2009 | Electrical apparatus for Potentially Explosive Atmospheres – General Requirements.(Harmonized) (IEC 60079-0:2007 Edition 5) |
| EN60079-11:2007 | Electrical apparatus for Potentially Explosive Atmospheres – Intrinsic Safety ‘i’.(Harmonized) (IEC 60079-11:2006 Ed 5) |
Pressure generation
PV 621, 622 abd 623 Specification
| Maximum pressure | PV 621 20 bar(300 psi) pneumatic PV 622 100 bar(1500 psi) pneumatic PV 623 1000 bar(15000 psi) hydraulic |
|---|---|
| Pressure media | PV 621 and PV 622 non-corrosive gases, PV 623 de-mineralized water or mineral oil(ISO viscosity grade < 22) |
| Operating temperature | -10° to 50°C (14° to 122°F) For water +4 ti +50°C()39 to 122°F |
| Storage temperature | -20 to 70°C (-4 to 158°F)(must be empty of water) |
| Shock and vibration | BS EN 61010:2001; Def stan 66-31, 8.4 cat III, 1m drop tested |
| Pressure safety | Pressure equipment derective class SEP |
| Size and weight | 450 mm x 280 mm x 235 mm, PV 621 2.65 kg, PV 622 3.30 kg, PV 623 3.75 kg |
PV 62X IS Pressure Station Specification
| Operation temperature | -10° to 40°C (14° to 104°F) Baseefa10ATEX0011X IECEx BAS 10.0003X Page 11 of 46 TRO753 Issue 1 EX II 2 G |
|---|---|
| EN60079-0:2009 | Electrical apparatus for Potentially Explosive Atmospheres - General Requirements.(Harmonized) (IEC 60079-0:2007 Edition 5) |
| EN60079-11:2007 | Electrical apparatus for Potentially Explosive Atmospheres - Intrinsic Safety 'i'.(Harmonized) (IEC 60079-11:2006 Ed 5) |
Download